Abstract

Background

Proton therapy relies on protons stopping at the depth specified in the treatment plan. Secondary photons and neutrons produced along the beam path can escape the target, providing a way to check the beam range without placing a detector inside the patient.

Approach

Monte Carlo simulations examined how a digital tracking calorimeter responds to secondary photons and neutrons. To test the simulated response, we placed a smaller silicon pixel telescope behind a water phantom and irradiated it with 100–200 MeV proton beams at the Danish Centre for Particle Therapy in Aarhus, Denmark.

Results

The simulations showed that aluminum between the sensor layers converts secondary photons and neutrons into charged particles the sensors can detect. The measured signal correlated with the nominal beam energy. Using only the number of detected hit clusters per primary proton, a simulation-based model with linear calibration estimated the beam energy with a root mean squared error of 1.34 MeV. Beam energy served as a proxy for proton range in the homogeneous water phantom.

Significance

The experiment shows that a calorimeter designed for proton computed tomography could also support range verification without a separate detector. The study demonstrates feasibility in a water phantom, not clinical performance.

Key Contributions

  • Experimental evidence that a compact silicon pixel telescope can detect a signal correlated with proton beam energy
  • Simulation analysis of how aluminum absorbers make secondary photons and neutrons detectable
  • Beam-energy estimation from a single detector feature: hit clusters per primary proton

Technologies & Methods

  • GATE/Geant4 Monte Carlo simulation of particle transport and detector response
  • ALPIDE silicon pixel sensors in a prototype digital tracking calorimeter
  • Water-phantom measurements with a medical proton beam